کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1247241 | 969813 | 2007 | 4 صفحه PDF | دانلود رایگان |
For the simple and fast preparation of highly reliable standard materials, a post-column reaction GC/FID system was developed and evaluated on the mixture of oxygen-containing organic compounds. The oxygen-containing organic compounds mixing solution were determined with the post-column reaction GC/FID system using n-dodecane as an internal calibration standard. Required value of relative expanded uncertainty as an original source of SI-traceable standard materials was within 1% and it aimed at this value as accuracy of the quantitative analysis. The results showed good agreement between the prepared concentrations and analytical values using post-column reaction GC/FID system. These results indicated that the post-column reaction GC/FID system would be used for getting SI-traceable values.
Journal: Talanta - Volume 72, Issue 5, 31 July 2007, Pages 1655–1658