The loading of multi-walled carbon nanotubes (MWNTs) and glucose oxidase (GOx) in the alternate layers of a glucose biosensor was first optimized based on a layer-by-layer construction on the surface of a graphite disk electrode. With the increasing of MWNTs/GOx layers, the response current to glucose was changed regularly and the response current reached a maximum value when the number of MWNTs/GOx layers was 6. Owing to a good electrical conductivity, strong adsorption and excellent bioconsistency of MWNTs, the (MWNTs/GOx)6 films-coated glucose biosensor had an excellent electrochemical properties. The response current of the (MWNTs/GOx)6 films-coated biosensor to 3 × 10− 2 M glucose was 1.63 μA while the response time was only 6.7 s. The linear range and the lowest detectable concentration of this biosensor was 5 × 10− 4∼1.5 × 10− 2 M and 0.9 × 10− 4 M, respectively.
Journal: Materials Science and Engineering: C - Volume 26, Issue 1, January 2006, Pages 113–117