کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1761602 | 1019655 | 2011 | 8 صفحه PDF | دانلود رایگان |
عنوان انگلیسی مقاله ISI
Four-Year Experience with a Clinical Ultrasound Quality Control Program
دانلود مقاله + سفارش ترجمه
دانلود مقاله ISI انگلیسی
رایگان برای ایرانیان
کلمات کلیدی
موضوعات مرتبط
مهندسی و علوم پایه
فیزیک و نجوم
آکوستیک و فرا صوت
پیش نمایش صفحه اول مقاله

چکیده انگلیسی
Ultrasound (US) quality control (QC) program data over a 4-year period from more than 45 scanners and more than 265 transducers were reviewed to optimize the program in terms of efficiency and effectiveness. Our program included evaluations of mechanical integrity, image uniformity, distance measurement accuracy and maximum depth of penetration (DOP). We computed failure rates and fraction of failures detected by each test. A total of 187 equipment problems were identified. Average annual scanner component and transducer failure rates were 10.5% and 13.9%, respectively. The mechanical integrity and uniformity evaluations detected 25.1% and 66.3% of all failures, respectively. Those evaluations plus defects detected by sonographers accounted for 98.4% of all detected failures. DOP and distance measurement accuracy were not effective at detecting equipment failures. For routine US QC, we recommend quarterly mechanical integrity and uniformity assessments of all transducers. A scanner with five transducers could be tested in an estimated 30 min or less.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Ultrasound in Medicine & Biology - Volume 37, Issue 8, August 2011, Pages 1350-1357
Journal: Ultrasound in Medicine & Biology - Volume 37, Issue 8, August 2011, Pages 1350-1357
نویسندگان
Nicholas J. Hangiandreou, Scott F. Stekel, Donald J. Tradup, Krzysztof R. Gorny, Deirdre M. King,