کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1762565 1019700 2007 10 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Finite Amplitude Measurements of the Nonlinear Parameter B/A for Liquid Mixtures Spanning a Range Relevant to Tissue Harmonic Mode
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم آکوستیک و فرا صوت
پیش نمایش صفحه اول مقاله
Finite Amplitude Measurements of the Nonlinear Parameter B/A for Liquid Mixtures Spanning a Range Relevant to Tissue Harmonic Mode
چکیده انگلیسی
The objective of this investigation was to measure the nonlinear parameter B/A using an enhanced finite amplitude distortion technique, based on nonlinear propagation effects analogous to those associated with tissue harmonic imaging. These measurements validate an improved method for measuring the nonlinear parameter B/A, the small-signal speed of sound, and the attenuation coefficient from a single set of ultrasonic measurements. To test the method, measurements were performed on 11 different mixtures of isopropyl alcohol (isopropanol) and water that span the range of concentrations from 0% to 100% isopropanol. Results for B/A ranging from approximately five to 11 were found to be reproducible and in good agreement with previously published values obtained using a thermodynamic method. (E-mail: james.g.miller@wustl.edu)
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Ultrasound in Medicine & Biology - Volume 33, Issue 4, April 2007, Pages 620-629
نویسندگان
, , , ,