کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1762738 1019713 2008 14 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
An Exposimetry System Using Tissue-Mimicking Liquid
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم آکوستیک و فرا صوت
پیش نمایش صفحه اول مقاله
An Exposimetry System Using Tissue-Mimicking Liquid
چکیده انگلیسی
Acoustic output measurements of diagnostic ultrasound scanners are currently performed in water and derated to approximate in situ values. The derating scheme ignores nonlinear propagation of sound waves and has been shown in previous numerical and experimental studies to tend to underestimate relevant pressure and intensity values in tissue mimicking media. This work describes an alternative method, which uses a tissue-mimicking liquid with attenuation coefficient slope of 0.3 dB/cm/MHz, speed of sound of 1540 m/s and nonlinearity parameter B/A of 7.5. The acoustic properties of this liquid are stable for at least 2 y after production. Initial results using a single M-mode configuration are presented. These results confirm that derating can significantly underestimate the pulse intensity integral and peak rarefactional pressure. (E-mail: tastiles@wisc.edu)
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Ultrasound in Medicine & Biology - Volume 34, Issue 1, January 2008, Pages 123-136
نویسندگان
, , ,