Performance measurement of the scintillator with optical fiber detector for boron neutron capture therapy
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Performance measurement of the scintillator with optical fiber detector for boron neutron capture therapy
چکیده انگلیسی

The thermal neutron flux can be easily measured in real time by using the scintillator with optical fiber (SOF) detector. However the irradiation damage under high-intensity neutron flux causes deterioration of the SOF detector due to radiation damage to the plastic scintillator in which 6LiF is blended. After irradiating the SOF detector for 4 h (thermal neutron fluence is approximately 2.0×1013 neutrons/cm2), the sensitivity of the SOF detector decreased by 3.0%. After irradiating the SOF detector for 2 months (thermal neutron fluence approximately 6.4×1014 neutrons/cm2), the sensitivity was reduced to 42% of baseline. Supposing that the thermal neutron fluence is 2×1012 neutrons/cm2 on the surface of a patient in a BNCT treatment, the sensitivity of the SOF detector is reduced by approximately 0.3%. This report presents investigations on the deterioration of the SOF detector in irradiation experiments.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Radiation and Isotopes - Volume 67, Issues 7–8, Supplement, July 2009, Pages S254–S257