Various quality control (QC) and quality assurance (QA) procedures of the boron neutron capture therapy (BNCT) beam at the Tsing Hua Open-pool Reactor (THOR) are established to ensure beam availability and quality. The QC/QA methods mainly employ foil activation and paired ionization chambers, respectively, for beam intensity check and dose assessment. Beam intensity is monitored on-line by using three dead-time corrected fission chambers. In addition to the periodic QC/QA activities regarding beam quality and the monitoring system, the quick QC/QA performed in an all-in-one phantom will be executed less than 70 min before the clinical treatment to guarantee beam quality. The QC/QA procedures have been gradually established and the actual performance satisfied the preset criteria defined for the BNCT facility at THOR.