Are Lateral Electronic Portal Images Adequate for Accurate On-Line Daily Targeting of the Prostate? Results of a Prospective Study
دانلود مقاله + سفارش ترجمه
دانلود مقاله ISI انگلیسی
رایگان برای ایرانیان
کلمات کلیدی
موضوعات مرتبط
مهندسی و علوم پایه
فیزیک و نجوم
تشعشع
پیش نمایش صفحه اول مقاله

چکیده انگلیسی
The purpose of this report was to evaluate the magnitude of the error that would be introduced if only a lateral (LAT) portal image, as opposed to a pair of orthogonal images, was used to verify and correct daily setup errors and organ motion in external beam radiation therapy (EBRT) of prostate cancer. The 3-dimensional (3D) coordinates of gold markers from 12 consecutive prostate patients were reconstructed using a pair of orthogonal images. The data were re-analyzed using only the LAT images. Couch moves from the 2-dimensional (2D)-only data were compared with the complete 3D data set. The 2D-only data provided couch moves that differed on average from the 3D data by 2.3 ± 3.0, 0.0 ± 0.0, and 0.8 ± 1.0 mm in the Lat, AP, and SI directions, respectively. Along AP and SI axes, the LAT image provided positional information similar to the orthogonal pair. The error along the LAT axis may be acceptable provided lateral margins are large enough. A LAT-only setup protocol reduces patient treatment times and increases patient throughput. In most circumstances, with exceptions such as morbidly obese patients, acquisition of only a LAT image for daily targeting of the prostate will provide adequate positional precision.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Medical Dosimetry - Volume 33, Issue 1, Spring 2008, Pages 22-29
Journal: Medical Dosimetry - Volume 33, Issue 1, Spring 2008, Pages 22-29
نویسندگان
Michael W. M.S., Damon B.S., Michele M.S.E.E., Andrea Ph.D., Lynn Ph.D., James M. Ph.D., Roger Ph.D., Jean Ph.D., Mack Ph.D.,