کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1880523 | 1043042 | 2013 | 5 صفحه PDF | دانلود رایگان |
عنوان انگلیسی مقاله ISI
Radiation leakage study for the Valencia applicators
دانلود مقاله + سفارش ترجمه
دانلود مقاله ISI انگلیسی
رایگان برای ایرانیان
موضوعات مرتبط
مهندسی و علوم پایه
فیزیک و نجوم
تشعشع
پیش نمایش صفحه اول مقاله

چکیده انگلیسی
Radiation leakage of up to 170% of the prescribed dose has been found at the back surface of these applicators. Although this side is not usually directed to the patient, in some applications such as the treatment of a lesion on the nose, special care must be taken to avoid unexpected and unnecessary irradiation of the eyes. A possible solution could be to add additional shielding to the applicator in order to reduce this leakage or to put some shielding to protect the eyes. Additionally, a new concept design of the Valencia applicators using more shielding material in the applicator backside is proposed.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Physica Medica - Volume 29, Issue 1, January 2013, Pages 60-64
Journal: Physica Medica - Volume 29, Issue 1, January 2013, Pages 60-64
نویسندگان
D. Granero, J. Perez-Calatayud, F. Ballester, Z. Ouhib,