Modeling of couch transmission in the RayStation treatment planning system
چکیده انگلیسی


• RayStation can reproduce measured couch attenuations within an uncertainty of 1.0%.
• Using a couch CT scan can be sub-optimal compared to a geometrical-model.
• Gamma pass-rates increase if the couch is considered in pre-treatment QA.
• Ignoring the couch in the TPS leads to notably underdose for clinical targets.
• The 3%/3 mm γ-criterion hardly detects the inaccuracy of ignoring the couch.

PurposeTo present our methods and results regarding the modeling of a carbon fiber couch (Varian Exact IGRT) in the RayStation treatment planning system (TPS).MethodsThree geometrical-models (GMs) were implemented in the TPS to represent the three different regions of the couch (thick, medium and thin). The materials and densities of each GM component were tuned to maximize the agreement between measured and calculated attenuations. Moreover, a couch computed-tomography (CT) scan was acquired and dosimetrically compared with the GMs. For validation, plan-specific quality assurance (QA) of VMAT plans (TG-119 cases, 5 prostate and 5 H&N clinical cases) was performed by comparing measured dose distributions with doses computed with and without including the GMs in the TPS.ResultsCouch attenuations up to 4.3% were measured (energy: 6MV). Compared to couch CT, GMs could be modified to optimize the agreement with measurements and reduce dependence on the dose grid resolution. For both couch CT and GM, absolute deviations between measured and calculated attenuations were within 1.0%. When including the GMs in plan-specific QA, global 2%/2 mm γ-pass rates showed an average improvement of 4.8% (p-value < 0.001, max +18.6%). The couch reduced the mean dose to targets by up to 2.4% of the prescribed dose for prostate cases and up to 1.4% for H&N cases.ConclusionsRayStation accurately considers the implemented couch GMs replicating measured attenuations within an uncertainty of 1.0%. Materials and densities are proposed for the Varian Exact IGRT couch. The results obtained justify introducing couch GMs in clinical routine.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Physica Medica - Volume 32, Issue 5, May 2016, Pages 735–740
نویسندگان
, , , , ,