Availability of applying diaphragm matching with the breath-holding technique in stereotactic body radiation therapy for liver tumors
ترجمه فارسی عنوان
در دسترس بودن تطبیق دیافراگم کاربردی با روش نفوذ در روش پرتودرمانی بدن سترونوتاکتیک برای تومورهای کبدی
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
چکیده انگلیسی


• We investigated the feasibility of diaphragm matching for liver tumors.
• Diaphragm matching suppresses variations in respiratory tumor motion.
• Diaphragm matching was useful in all (C-C, A-P, L-R) direction.
• Although it was not showed in our previous study, PTV margin could be reduced.

PurposeImage-guided radiotherapy (IGRT) based on bone matching can produce large target-positioning errors because of expiration breath-hold reproducibility during stereotactic body radiation therapy (SBRT) for liver tumors. Therefore, the feasibility of diaphragm-based 3D image matching between planning computed tomography (CT) and pretreatment cone-beam CT was investigated.MethodsIn 59 liver SBRT cases, Lipiodol uptake after transarterial chemoembolization was defined as a tumor marker. Further, the relative isocenter coordinate that was obtained by Lipiodol matching was defined as the reference coordinate. The distance between the relative isocenter coordinate and reference coordinate, which was obtained from diaphragm matching and bone matching techniques, was defined as the target positioning error. Furthermore, the target positioning error between liver matching and Lipiodol matching was evaluated.ResultsThe positioning errors in all directions by the diaphragm matching were significantly smaller than those obtained by using by the bone matching technique (p < 0.05). Further, the positioning errors in the A-P and C-C directions that were obtained by using liver matching were significantly smaller than those obtained by using bone matching (p < 0.05). The estimated PTV margins calculated by the formula proposed by van Herk for diaphragm matching, liver matching, and bone matching were 5.0 mm, 5.0 mm, and 11.6 mm in the C-C direction; 3.6 mm, 2.4 mm, and 6.9 mm in the A-P direction; and 2.6 mm, 4.1 mm, and 4.6 mm in the L-R direction, respectively.ConclusionsDiaphragm matching-based IGRT may be an alternative image matching technique for determining liver tumor positions in patients.

Graphical AbstractFigure optionsDownload high-quality image (66 K)Download as PowerPoint slide

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Physica Medica - Volume 32, Issue 4, April 2016, Pages 557–561
نویسندگان
, , , , , , , , , , , ,