Commissioning and initial experience with a commercial software for in vivo volumetric dosimetry
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Commissioning and initial experience with a commercial software for in vivo volumetric dosimetry
چکیده انگلیسی


• Commissioning of Dosimetry Check software for VMAT in vivo dosimetry is presented.
• Simple and clinical plans are created for homogeneous and heterogeneous phantoms.
• Dose differences and gamma analysis between Dosimetry Check and TPS are studied.
• Total dose differences at isocentre are less than 2% for homogeneous phantoms.
• Discrepancies reach up to 15% in the presence of heterogeneities.

Introduction and purposeDosimetry Check (DC) (Math Resolutions) is a commercial EPID-based dosimetry software, which allows performing pre-treatment and transit dosimetry. DC provides an independent verification of the treatment, being potentially of great interest due to the high benefits of the in vivo volumetric dosimetry, which guarantee the treatment delivery and anatomy constancy. The aim of this work is to study the differences in dose between DC and the Treatment Planning System (TPS) to establish an accuracy level of the system.Material and methodsDC v.3.8 was used along with Varian Clinac iX accelerator equipped with EPID aS1000 and Eclipse v.10.0 with AAA and Acuros XB calculation algorithms. The DC evaluated version is based on a pencil beam calculation algorithm. Various plans were generated over several homogeneous and heterogeneous phantoms. Isocentre point doses and gamma analysis were evaluated.ResultsTotal dose differences at the isocentre between DC and TPS for the studied plans are less than 2%, but single field contributions achieve greater values. In the presence of heterogeneities, the discrepancies can reach up to 15%. In transit mode, DC does not consider properly the couch attenuation, especially when there is an air gap between phantom and couch.ConclusionsThe possibility of this in vivo evaluation and the potentiality of this new system have a very positive impact on improving patient QA. But improvements are required in both calculation algorithm and integration with the record and verify system.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Physica Medica - Volume 30, Issue 8, December 2014, Pages 954–959
نویسندگان
, , , , , , , , ,