کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1882625 | 1043244 | 2013 | 8 صفحه PDF | دانلود رایگان |

The beam quality correction factor kQ,Q0kQ,Q0 and the perturbation factor pQ , commonly considered in dosimetry with ionization chambers, were calculated for the NE2571 and the Standard Imaging A19 and A12S chambers, using the Monte Carlo simulation code PENELOPE. For the NE2571 chamber, the values of kQ,Q0kQ,Q0 obtained are in very good agreement with those found in previous works by Wulff et al. and Muir and Rogers with the code EGSnrc and also with the experimental results summarized in the NCS code of practice. For pQ, a difference of ∼0.4% has been found between our results and those obtained with EGSnrc for 60Co and this difference increases slightly with TPR1020 values. These factors have been calculated also for the A19 and A12S chambers of Standard Imaging. The values of kQ,Q0kQ,Q0 show reasonable agreement with those recently calculated by Muir and Rogers and the measurements of McEwen.
Journal: Physica Medica - Volume 29, Issue 2, March 2013, Pages 163–170