Influence of different Fletcher-Suit applicator geometries on sagittal dose distribution
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Influence of different Fletcher-Suit applicator geometries on sagittal dose distribution
چکیده انگلیسی
Strong correlation was found between the ovoid separation and dose values to applicator points defined in sagittal direction of FS applicator arrangements. Also strong correlation was between the ovoids' sagittal position with respect to the tandem and the applicator point defined towards the rectum, while the ovoid symmetry had no influence on the sagittal dose distribution. The standard deviations of inter-fractional variation of the ovoid separation and the ovoids' sagittal position were within ±5.2 mm and ±10.2 mm respectively. The inter-fractional variations in FS applicator geometry resulted in variation in dose levels at the applicator points ±0.8 Gy typically, while the largest value was ±1.6 Gy.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Physica Medica - Volume 26, Issue 1, January 2010, Pages 49-54