A comparative contrast perception phantom image of brain CT study between high-grade and low-grade liquid crystal displays (LCDs) in electronic medical charts
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
A comparative contrast perception phantom image of brain CT study between high-grade and low-grade liquid crystal displays (LCDs) in electronic medical charts
چکیده انگلیسی

The purpose of this study was to clarify whether non-medical-grade liquid crystal displays (LCDs) are acceptable for the soft-copy reading of brain CTs. Four kinds of color LCDs with different image quality levels were used: medical-grade LCD, low-grade general LCD calibrated with the grayscale display function (GSDF), low-grade general LCD calibrated with gamma 2.2 and a notebook personal computer display panel. In Osirix’s standard window setting for brain CTs, the average CT values of brain parenchyma in 100 cases were correlated with a grayscale level ranging from 71 to 91 in a 256-step grayscale. At these gray levels, the image contrast on the two low-grade LCDs calibrated with gamma 2.2 was higher than that on the medical-grade LCD. Eleven healthy volunteers participated in the contrast perception study, which used electronically generated target phantom images that simulated subtle abnormalities with a low or high attenuation difference in brain parenchyma. The three low-grade LCDs showed correct response rates and reaction times that were superior to those of the medical-grade display. The grayscale calibrations, GSDF or gamma 2.2, are likely to be more critical than the display grade, suggesting that the use of a low-grade LCD may be acceptable in the image contrast of brain CT.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Physica Medica - Volume 27, Issue 2, April 2011, Pages 109–116
نویسندگان
, , , , ,