• The IBA Razor overresponds in small fields as with other silicon diodes.
• Correction factors for the overresponse are presented for Varian linacs.
• The correction factor of FF beams was compared to that of FFF beams.
• The overresponse can be predicted mathematically based on the field size.
The IBA Razor diode supersedes the IBA SFD and is intended for use in small fields. However, its behaviour in small fields has not yet been quantified. In this work, we examine the response of the Razor diode against the air core scintillation dosimeter (FOD) and Gafchromic film in photon beams from three Varian linac beams. Fields between 4 mm and 30 mm in width were measured, both with and without a flattening filter and at two energies. The Razor exhibited an over-response of up to 4.5% for MLC collimated fields and 7.1% for stereotactic cone collimated fields. The presence of the flattening filter altered the over-response by up to 1.5%. The small field correction factors are tabulated and agree with the mathematical relation of Liu et al. (2014). Four samples of the Razor were used, two having received a significant prior dose. The correction factors for the four samples differed and may depend on their dose history.
Journal: Physica Medica - Volume 32, Issue 8, August 2016, Pages 1025–1029