کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
193872 459779 2008 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
An ECL biosensor for glucose based on carbon-nanotube/Nafion film modified glass carbon electrode
موضوعات مرتبط
مهندسی و علوم پایه مهندسی شیمی مهندسی شیمی (عمومی)
پیش نمایش صفحه اول مقاله
An ECL biosensor for glucose based on carbon-nanotube/Nafion film modified glass carbon electrode
چکیده انگلیسی

Glucose oxidase was encapsulated in carbon-nanotube/Nafion film modified glass carbon electrode and was used as electrochemiluminescence (ECL) biosensor for glucose. The glucose oxidase can be fixed firmly in the Nafion film and carbon nanotubes offer excellent electrocatalytic activity toward luminol and hydrogen peroxide liberated in enzymatic reaction between glucose oxidase and glucose, which would enable sensitive determination of glucose. Under the optimum condition, the linear response range of glucose was found to be 5.0 × 10−6 to 8.0 × 10−4 mol/L, and the detection limit (defined as the concentration that could be detected at the signal-to-noise ratio of 3) was 2.0 × 10−6 mol/L. The present carbon-nanotube/Nafion biocomposite glucose oxidase ECL biosensor showed excellent properties for sensitive determination for glucose with good reproducibility and stability, and it has been used to determine the glucose concentrations in real serum samples with the satisfactory results.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Electrochimica Acta - Volume 53, Issue 5, 1 January 2008, Pages 2396–2401
نویسندگان
, , ,