کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
195318 459812 2007 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Low potential detection of glucose at carbon nanotube modified glassy carbon electrode with electropolymerized poly(toluidine blue O) film
موضوعات مرتبط
مهندسی و علوم پایه مهندسی شیمی مهندسی شیمی (عمومی)
پیش نمایش صفحه اول مقاله
Low potential detection of glucose at carbon nanotube modified glassy carbon electrode with electropolymerized poly(toluidine blue O) film
چکیده انگلیسی

A mediator glucose biosensor has been constructed by immobilizing glucose oxidase at electropolymerized poly(toluidine blue O) film on carbon nanotube modified glass carbon electrode. The toluidine blue O moieties served as redox mediators for enzymatic glucose oxidation and as polymeric network to maintain the biosensor activity. Great enhancement in current response was observed for the glucose biosensor. The detection potential could be decreased to −0.1 V (versus Ag|AgCl), where common interferences such as ascorbic acid, uric acid and acetamidophenol were not oxidized to cause interferences. The amperometric glucose biosensor offered a sensitivity of 14.5 mA M−1 cm−2 for the linear range of 1–7 mM.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Electrochimica Acta - Volume 53, Issue 2, 1 December 2007, Pages 278–284
نویسندگان
, ,