The adenosine triphosphate test is a rapid and reliable audit tool to assess manual cleaning adequacy of flexible endoscope channels
موضوعات مرتبط
علوم زیستی و بیوفناوری ایمنی شناسی و میکروب شناسی میکروب شناسی
پیش نمایش صفحه اول مقاله
The adenosine triphosphate test is a rapid and reliable audit tool to assess manual cleaning adequacy of flexible endoscope channels
چکیده انگلیسی

BackgroundThe study objective was to verify that the adenosine triphosphate (ATP) benchmark of <200 relative light units (RLUs) was achievable in a busy endoscopy clinic that followed the manufacturer's manual cleaning instructions.MethodsAll channels from patient-used colonoscopes (20) and duodenoscopes (20) in a tertiary care hospital endoscopy clinic were sampled after manual cleaning and tested for residual ATP. The ATP test benchmark for adequate manual cleaning was set at <200 RLUs. The benchmark for protein was <6.4 μg/cm2, and, for bioburden, it was <4-log10 colony-forming units/cm2.ResultsOur data demonstrated that 96% (115/120) of channels from 20 colonoscopes and 20 duodenoscopes evaluated met the ATP benchmark of <200 RLUs. The 5 channels that exceeded 200 RLUs were all elevator guide-wire channels. All 120 of the manually cleaned endoscopes tested had protein and bioburden levels that were compliant with accepted benchmarks for manual cleaning for suction-biopsy, air-water, and auxiliary water channels.ConclusionOur data confirmed that, by following the endoscope manufacturer's manual cleaning recommendations, 96% of channels in gastrointestinal endoscopes would have <200 RLUs for the ATP test kit evaluated and would meet the accepted clean benchmarks for protein and bioburden.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: American Journal of Infection Control - Volume 41, Issue 3, March 2013, Pages 249–253
نویسندگان
, , ,