کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
2726533 1146489 2006 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
The importance of clinical based quality assurance (QA) on an RT250 Phillips deep X-ray unit
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی رادیولوژی و تصویربرداری
پیش نمایش صفحه اول مقاله
The importance of clinical based quality assurance (QA) on an RT250 Phillips deep X-ray unit
چکیده انگلیسی

Experimentation on an RT250 Phillips X-ray machine, used to treat superficial skin lesions, led to the discovery of deformed treatment fields, which was investigated using X-ray film. The cause of the field irregularity was found to be due to the circular applicators becoming misaligned in the housing. The treatment field was deformed when using the 3 cm and 3.5 cm applicators giving grounds to recommend the removal of these applicators from clinical use.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Radiography - Volume 12, Issue 4, November 2006, Pages 314–317