The sensitivity of direct identification from positive BacT/ALERT™ (bioMérieux) blood culture bottles by matrix-assisted laser desorption ionization time-of-flight mass spectrometry is low
موضوعات مرتبط
علوم زیستی و بیوفناوری ایمنی شناسی و میکروب شناسی میکروب شناسی
پیش نمایش صفحه اول مقاله
The sensitivity of direct identification from positive BacT/ALERT™ (bioMérieux) blood culture bottles by matrix-assisted laser desorption ionization time-of-flight mass spectrometry is low
چکیده انگلیسی

Recently, matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF MS) has been presented as a novel method for the direct identification of bacteria from positive blood culture bottles. The rate of the MALDI TOF MS-based identification in the present study from positive BacT/ALERT (bioMérieux, Marcy l'Etoile, France) blood culture bottles was 30%, which is far below the previously reported sensitivities using the BACTEC (Becton Dickinson, Franklin Lakes, NJ, USA) system. We also found evidence that the Biotyper algorithm did not identify a second pathogen in cases of positive BacT/ALERT blood culture bottles containing two different species.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Clinical Microbiology and Infection - Volume 17, Issue 2, February 2011, Pages 192–195