Liikavainio T, Bragge T, Hakkarainen M, Jurvelin JS, Karjalainen PA, Arokoski JP. Reproducibility of loading measurements with skin-mounted accelerometers during walking.ObjectiveTo examine reproducibility of load measurements with skin-mounted accelerometers (SMAs) during walking.DesignReliability study.SettingA motion analysis laboratory.ParticipantsTen healthy young men.InterventionsNot applicable.Main Outcome MeasuresTwo triaxial accelerometers were fixed to the subjects’ skin above and below the knee joint. The subjects walked barefoot at their preferred speed and at a constant speed (1.3m/s, ±5%) in a gait laboratory and along a corridor. The same protocol was repeated over 2 days. Initial peak acceleration (IPA), peak-to-peak (PP) acceleration, and maximal and average acceleration transient rates (ATRs) were calculated. The coefficient of variation (CV) and Pearson linear correlation coefficient were calculated to measure reproducibility of SMA load measurements.ResultsIPA and PP acceleration had good interday repeatability (CV <15%). The repeatability of average ATR and maximal ATR parameters was generally not acceptable. The loading variables obtained from ground reaction forces and SMA measurements during gait revealed high linear correlations, indicating that with SMA measurements it is possible to predict certain ground reaction force loading parameters.ConclusionsSMAs are practical for use in clinical environments to collect acceleration data that may be used to estimate joint loads.
Journal: Archives of Physical Medicine and Rehabilitation - Volume 88, Issue 7, July 2007, Pages 907–915