کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
5031459 1369950 2017 17 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Using silver nanoparticle and thiol graphene quantum dots nanocomposite as a substratum to load antibody for detection of hepatitis C virus core antigen: Electrochemical oxidation of riboflavin was used as redox probe
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Using silver nanoparticle and thiol graphene quantum dots nanocomposite as a substratum to load antibody for detection of hepatitis C virus core antigen: Electrochemical oxidation of riboflavin was used as redox probe
چکیده انگلیسی
Compared to the other redox probes, riboflavin is superior in its oxidization in negative potential range, where the number of interfering species for riboflavin is much fewer. The proposed immunosensor showed wide linear range from 0.05 pg mL−1 to 60 ng mL−1 with limit of detection of 3 fg mL−1. This novel immunosensor was used to analyze the serum sample. The immunosensor provides a convenient, low-cost and simple method for HCV core antigen detection and proposes new horizons for quantitative detection of antigen in the clinical diagnosis.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Biosensors and Bioelectronics - Volume 89, Part 2, 15 March 2017, Pages 946-951