High-resolution two-dimensional liquid chromatography analysis of key linker drug intermediate used in antibody drug conjugates
چکیده انگلیسی


- Over 15 residual co-eluting impurities resolved from the main component on a complementary secondary column using 2D-LC.
- Over 5 fold enhancement in secondary column sensitivity with detection limits < 0.01% suitable for ultra-trace analysis.
- Assess overall purity of main component/s of interest based on primary and secondary column purity.
- The 2D-LC is highly reproducible and comparable to conventional HPLC.
- 2D-LC analysis of linker drug intermediate enabling better understanding of impurities with downstream impact to ADCs.

In this manuscript, the application of high-resolution sampling (HRS) two-dimensional liquid chromatography (2D-LC) in the detailed analysis of key linker drug intermediate is presented. Using HRS, selected regions of the primary column eluent were transferred to a secondary column with fidelity enabling qualitative and quantitative analysis of linker drugs. The primary column purity of linker drug intermediate ranged from 88.9% to 94.5% and the secondary column purity ranged from 99.6% to 99.9%, showing lot-to-lot variability, significant differences between the three lots, and substantiating the synthetic and analytical challenges of ADCs. Over 15 impurities co-eluting with the linker drug intermediate in the primary dimension were resolved in the secondary dimension. The concentrations of most of these impurities were over three orders of magnitude lower than the linker drug. Effective peak focusing and high-speed secondary column analysis resulted in sharp peaks in the secondary dimension, improving the signal-to-noise ratios. The sensitivity of 2D-LC separation was over five fold better than conventional HPLC separation. The limit of quantitation (LOQ) was less than 0.01%. Many peaks originating from primary dimension were resolved into multiple components in the complementary secondary dimension, demonstrating the complexity of these samples. The 2D-LC was highly reproducible, showing good precision between runs with%RSD of peak areas less than 0.1 for the main component. The absolute difference in the peak areas of impurities less than 0.1% were within ±0.01% and for impurities in the range of 0.1%-0.3%, the absolute difference were ±0.02%, which are comparable to 1D-LC. The overall purity of the linker drug intermediate was determined from the product of primary and secondary column purity (HPLC Purity = %peak area of main component in the primary dimension × %peak area of main component in the secondary dimension). Additionally, the 2D-LC separation enables the determination of potential impurities that could impact the downstream process, like ADCs stability, efficacy and patient safety. Peak capacity of this magnitude, sensitivity and reproducibility of 2D-LC for resolving structurally similar impurities co-eluting with the main component has not been demonstrated to date. This application clearly demonstrates the power of 2D-LC in detailed analysis of structurally similar, co-eluting impurities from key linker drug intermediate used in ADCs that is impossible to achieve by conventional 1D-LC.