- A new approach to calibrate SEC-light scattering setups is proposed.
- The approach requires neither dn/dc of calibrant nor of analyte.
- Only molar mass of the reference standard and the injected masses of calibrant and analyte are needed.
A new approach for the calibration of SEC-light scattering (SEC-LS) setups is proposed, which requires solely the molar mass of a reference polymer. Neither the specific refractive index increment of the calibrant nor of the analyte is required. Comparison of the molar masses derived in different solvents for a large number of chemically different polymers shows that the new approach yields the same molar masses as if molar masses were derived using dn/dc to calibrate the light scattering setup. The approach therefore allows easier determination of molar masses by SEC-LS.
Journal: Journal of Chromatography A - Volume 1520, 20 October 2017, Pages 103-106