Development of a split-flow system for high precision variable sample introduction in supercritical fluid chromatography
چکیده انگلیسی


- Split-flow sample injection is proposed for supercritical fluid chromatography.
- Differential pressure between two back pressure regulators splits the mobile phase.
- A make-up pump allows pressure control independent of the column pressure drop.
- This robust system was successfully applied to nine chemically diverse pesticides.

In this study, we propose a novel variable sample injection system based on full-loop injection, named the split-flow sample introduction system, for application in supercritical fluid chromatography (SFC). In this system, the mobile phase is split by the differential pressure between two back pressure regulators (BPRs) after full-loop injection suitable for SFC, and this differential pressure determines the introduction rate. Nine compounds with a wide range of characteristics were introduced with high reproducibility and universality, confirming that a robust variable sample injection system was achieved. We also investigated the control factors of our proposed system. Sample introduction was controlled by the ratio between the column-side pressure drops in splitless and split flow, ΔPcolumnsideinsplitless and ΔPcolumnsideinsplit, respectively, where ΔPcolumnsideinsplitless is related to the mobile phase flow rate and composition and the column resistance. When all other conditions are kept constant, increasing the make-up flow induces an additional pressure drop on the column side of the system, which leads to a reduced column-side flow rate, and hence decreased the amount of sample injected, even when the net pressure drop on the column side remains the same. Thus, sample introduction could be highly controlled at low sample introduction rate, regardless of the introduction conditions. This feature is advantageous because, as a control factor, the solvent in the make-up pump is independent of the column-side pressure drop.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Chromatography A - Volume 1515, 15 September 2017, Pages 218-231
نویسندگان
, , , , , ,