- A dynamic headspace needle-trap device was developed for the extraction of some VOCs.
- A newly prepared ZIF-8-NPC material was used as an efficient adsorbent for the NTD.
- The established method was applied to determine eleven VOCs in pen ink samples.
Some harmful volatile organic compounds (VOCs), such as methylbenzene, ethylbenzene, xylene, chlorobenzene and bromobenzene, are the commonly found chemicals in pen inks. In this work, a dynamic headspace needle-trap device (D-HS-NTD) with a ZIF-8-derived nanoporous carbon (ZIF-8-NPC) as the adsorbent was developed for the extraction of some VOCs in different pen inks prior to GC-MS detection. The main important influencing experimental parameters including the flow rate of the purge gas N2, extraction temperature, extraction time, desorption temperature and desorption time for the extraction were optimized to obtain a high extraction efficiency. Under the optimized conditions, a good linearity was obtained in the concentration range of 0.1-400 μg kgâ1 with the correlation coefficients (r) ranging from 0.9911 to 0.9990 for the eleven VOCs. The LODs at a signal-to-noise ratio of 3 (S/N = 3) were measured to be 10-20 ng kgâ1 for the VOCs. The developed method was applied to determine the VOCs from 20 pen inks. The recoveries of the VOCs for the method at the spiking levels of 0.5 and 20.0 μg kgâ1 fell in the range from 80.0% to 108%.
Journal: Journal of Chromatography A - Volume 1513, 1 September 2017, Pages 27-34