کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
5135323 | 1493442 | 2017 | 5 صفحه PDF | دانلود رایگان |

- Schlieren imaging microscopy is employed for moving boundary electrophoresis analyses.
- The multichannel chip with Schlieren detection enhanced the throughput and eliminated the lens effect.
- The effect of channel's geometry and dimension were investigated.
- The diffusion coefficients of sample proteins were determined under static and dynamic conditions.
The concentration gradient detection method based on the Schlieren optics employed for electrophoresis analyses by extending the technology to a multi-channel system using a prototyped microfluidic chip (thinXXS Micro-technology, Germany). The results prove that coupling a chip-based microfluidic device with Schlieren detection is an appropriate approach to improve the electrophoretic separations. The effects of channel's geometry and dimension were investigated by conducting the experiments in channels with different cross sectional areas. Fast kinetic data acquisition of the charge-coupled device (CCD) camera facilitated recording of a time sequence of optical images, demonstrating the potential of the CCD camera as a powerful tool for studying dynamic processes such as diffusion. Diffusion coefficients of sample proteins were measured under static and dynamic conditions, where the static mode demonstrated more accurate results. Furthermore, the Fourier transformation was employed to improve the Schlieren images for quantitative analysis of the diffusion coefficient measurement.
Journal: Journal of Chromatography A - Volume 1484, 10 February 2017, Pages 93-97