کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
5136232 | 1494001 | 2017 | 6 صفحه PDF | دانلود رایگان |
- New microextraction technique MIP-Tablet was developed.
- MIP-Tablet was applied for extraction of amphetamine in human urine samples.
- The validation showed that the method is selective and accurate.
- The MIP-Tablet could be used for more than 20 extractions.
Amphetamine selective molecularly imprinted sol-gel polymer tablets, MIP-tablets, for solid-phase microextraction of biofluid samples were prepared. An acetonitrile solution of deuterated amphetamine template and silane precursor, 3-(propylmethacrylate) trimethoxysilane, was soaked into the pores of polyethylene tablet substrates and polymerized by an acid-catalysed sol-gel process. Application of the resultant MIP-tablets to extract amphetamine from human urine samples followed by LC-MS/MS analysis was investigated. The extraction protocol was optimised with respect to pH of sample, addition of sodium chloride, extraction time, desorption solvent and desorption time. The final analysis method determined amphetamine in human urine with a limit of detection (LOD) of 1.0Â ng/mL and a lower limit of quantification (LLOQ) of 5Â ng/mL. Validation demonstrated accuracy of the method was 91.0-104.0% and inter-assay precision was 4.8-8.5% (RSD). Extraction recovery was 80%. The MIP-tablets could be re-used and the same tablet could be employed for more than twenty extractions.
Journal: Journal of Chromatography B - Volume 1063, 15 September 2017, Pages 130-135