کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
529 46 2013 8 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Strain-induced crack formations in PDMS/DXA drug collars
موضوعات مرتبط
مهندسی و علوم پایه مهندسی شیمی بیو مهندسی (مهندسی زیستی)
پیش نمایش صفحه اول مقاله
Strain-induced crack formations in PDMS/DXA drug collars
چکیده انگلیسی

Drug-eluting systems are currently used in cardiac leads in order to reduce inflammation and fibrosis at the lead–tissue interface. Drug release from these drug delivery systems can be modulated by the manufacturing processes used to create the drug systems and assemble them onto the cardiac lead. In this study, scanning electron microscopy, atomic force microscopy and Raman microscopy are employed to explore the material characteristics of a polydimethylsiloxane–dexamethasone acetate drug collar used on cardiac leads when varying the strain during collar assembly on the lead. A novel test fixture was created in order to investigate these drug collars under simulated stresses. Measurements of the collar while fitted to a rod revealed microcracks that are hypothesized to affect the drug release performance, resulting in increased drug elution. It was found that the strain that occurs during assembly of the collar onto the lead is a key factor in the formation of these microcracks. Results also suggest that cracks tend to form in areas of high drug particle density, and propagate between drug particles.

Figure optionsDownload high-quality image (186 K)Download as PowerPoint slide

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Acta Biomaterialia - Volume 9, Issue 7, July 2013, Pages 7335–7342
نویسندگان
, , , , , , ,