کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
5497925 1532072 2017 26 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Longitudinal monitoring of whole body counter NaI(Tl) detector efficiency
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Longitudinal monitoring of whole body counter NaI(Tl) detector efficiency
چکیده انگلیسی
Assessing accuracy of radiation counting systems over time is critical. We examined long-term WBC performance in detail. Efficiency factors for 54 detectors were updated annually over several years. Newer efficiency values were compared with baseline and with annual values. Overall system efficiency has declined (−1.9% over 3 yrs) and appears to be doing so at an increasing rate. Having more specific performance data on individual components can make the process of system maintenance and repair more straightforward and efficient.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Radiation and Isotopes - Volume 125, July 2017, Pages 74-79