Is a single isocenter sufficient for volumetric modulated arc therapy radiosurgery when multiple intracranial metastases are spatially dispersed?
ترجمه فارسی عنوان
آیا یک ایزو سنتر برای جراحی رادیواکتیوی مدولاسیون تابشی کافی است وقتی که متاستازهای متعدد داخل جمجمه فضایی پراکنده می شوند؟
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
چکیده انگلیسی
Previous work demonstrated improved dosimetry of single isocenter volumetric modulated arc therapy (VMAT) of multiple intracranial targets when they are located ≤ 4 cm from isocenter because of narrower multileaf collimators (MLCs). In follow-up, we sought to determine if decreasing isocenter-target distance (diso) by using 2 to 3 isocenters would improve dosimetry for spatially dispersed targets. We also investigated the effect of a maximum dose constraint during VMAT optimization, and the dosimetric effect of the number of VMAT arcs used for a larger number of targets (i.e., 7 to 9). We identified radiosurgery cases that had multiple intracranial targets with diso of at least 1 target > 5 cm. A single isocenter VMAT plan was created using a standardized 4-arc technique with 18 Gy per target. Each case was then replanned (1) using 2 to 3 isocenters, (2) including a maximum dose constraint per target, and in the case of 7 to 9 targets, (3) using 3 to 6 arcs. Dose evaluation included brain V6 Gy and V12 Gy, and conformity index (CI), gradient index (GI), and heterogeneity index (HI) per target. Two isocenters were sufficient to limit diso to ≤ 4 cm and ≤ 5 cm for 11/15 and 13/15 cases, respectively; after replanning with 2 to 3 isocenters, diso decreased from 5.8 ± 2.8 cm (2.3 14.9) to 2.5 ± 1.4 cm (0 5.2). All dose statistics improved on average, albeit modestly: V6 Gy = 6.9 ± 7.1%, V12 Gy = 0.9% ± 4.4%, CI = 2.6% ± 4.6%, GI = 0.9% ± 12.7%, and HI = 2.6% ± 5.2%; however, the number of arcs doubled and monitor units increase by nearly 2-fold. A maximum dose constraint had a negative effect on all dose indices, increasing V12 Gy by 9.7 ± 6.9%. For ≥ 7 targets, increasing number of arcs to > 3 improved CI, V12 Gy, and V6 Gy. A single isocenter is likely sufficient for VMAT radiosurgery of multiple intracranial metastases. Optimal treatment plan quality is achieved when no constraint is placed on the maximum target dose; for cases with many targets at least 4 arcs are needed for optimal plan quality.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Medical Dosimetry - Volume 41, Issue 4, Winter 2016, Pages 285-289
نویسندگان
, , , , , , ,