Technical noteImpact of deformable image registration accuracy on thoracic images with different regularization weight parameter settings
ترجمه فارسی عنوان
توجه فنی تاثیر دقت صحت ثبت نام تصویر بر روی تصاویر قفسه سینه با تنظیمات تنظیم پارامترهای مختلف وزن تنظیم
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
چکیده انگلیسی


- An investigation on improvement of DIR accuracy using a script is presented.
- The regularization weight has an influence on the DIR accuracy in the thoracic.
- A high regularization weight could be helpful in improving DIR accuracy.

PurposeWe assessed the deformable image registration (DIR) accuracy of thoracic images under different regularization weights using commercially available DIR software.MethodsThe thoracic 4-dimensional (4D) CT images of 10 patients were used. The datasets for these patients were provided by DIR-lab (www.dir-lab.com) and included a coordinate list of 300 anatomic landmarks that had been manually identified. The ANAtomically CONstrained Deformation Algorithm (ANACONDA) of RayStation (RaySearch Laboratories, Stockholm, Sweden) was used to deform the peak-inhale to peak-exhale images under different regularization weights (4, 40, 400-default setting, 1500, 4000, 10,000, 15,000, 20,000, 30,000, and 40,000). The regularization weights were changed using a script. The registration error (RE) was determined by calculating the difference at each landmark point between the displacement calculated by the DIR software and that calculated by the landmark. We measured the computation time for each regularization weight setting.ResultsHigh regularization weights resulted in a smaller RE than that observed with lower regularization weights. The RE decreases rapidly with increase in regularization weight before reaching a plateau. No significant difference was found between a regularization weight of 400 and regularization weights of 4, 40, 4000 or 40,000 (P value >0.05). The range of the average time was 8.4-12.2 s.ConclusionsWe concluded that the default setting for ANACONDA is stable with respect to regularization weight in the thoracic region.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Physica Medica - Volume 42, October 2017, Pages 108-111
نویسندگان
, , , , , , , ,