Comparative evaluation of a laboratory developed real-time PCR assay and the RealStar® HHV-6 PCR Kit for quantitative detection of human herpesvirus 6
موضوعات مرتبط
علوم زیستی و بیوفناوری ایمنی شناسی و میکروب شناسی ویروس شناسی
پیش نمایش صفحه اول مقاله
Comparative evaluation of a laboratory developed real-time PCR assay and the RealStar® HHV-6 PCR Kit for quantitative detection of human herpesvirus 6
چکیده انگلیسی


- A highly accurate quantitative real-time PCR assay for detecting HHV-6 in clinical specimens is described.
- The assay compares favorably to a commercially available HHV-6 qPCR kit.
- The assay is highly sensitive and specific for the detection of HHV-6 reactivation in hematopoietic stem cell transplant recipients.

BackgroundHHV-6 reactivation in immunocompromised patients is common and may be associated with serious morbidity and mortality; therefore, early detection and initiation of therapy might be of benefit. Real-time PCR assays allow for early identification of HHV-6 reactivation to assist in providing a timely response. Thus, we compared the performance of an in-house developed HHV-6 quantitative PCR assay with a commercially available kit, the RealStar® HHV-6 PCR Kit.MethodThe analytical sensitivity, analytical specificity, linearity, precision and accuracy of the in-house developed HHV-6 qPCR assay were evaluated. The diagnostic performance of the in-house HHV-6 qPCR assay was compared with the RealStar® HHV-6 PCR Kit, using 72 clinical specimens and 17 proficiency testing samples.ResultsLinear regression analysis of the quantitative results showed a dynamic range from 2 to 10 log10 copies/ml and a coefficient of determination (R2) of 0.999 for the in-house assay. A dilution series demonstrated a limit of detection and a limit of quantification of 1.7 log10 and 2 log10 copies/ml, respectively. The precision of the assay was highly reproducible among runs with coefficients of variance (CV) ranging from 0.27% to 4.37%. A comparison of 27 matched samples showed an excellent correlation between the quantitative viral loads measured by the in-house HHV-6 qPCR assay and the RealStar® HHV-6 PCR Kit (R2 = 0.926; P < 0.0001), with an average bias of −0.24 log10 copies/ml.ConclusionsThe in-house developed HHV-6 qPCR method is a sensitive and reliable assay with lower cost for the detection and quantification of HHV-6 DNA when compared to the RealStar® HHV-6 PCR Kit.