Quantifying the effects of operational parameters on the counting efficiency of a condensation particle counter using response surface Design of Experiments (DoE)
موضوعات مرتبط
مهندسی و علوم پایه علوم زمین و سیارات علم هواشناسی
پیش نمایش صفحه اول مقاله
Quantifying the effects of operational parameters on the counting efficiency of a condensation particle counter using response surface Design of Experiments (DoE)
چکیده انگلیسی
The size-resolved counting efficiency of an in-house ultrafine condensation particle counter (IUCPC) being challenged by sodium chloride (NaCl) particles was experimentally determined with a reference electrometer TSI 3068B. The influence of several operational parameters including the saturator temperature (35 to 45 ℃), the condenser temperature (10 to 25 ℃), the capillary sample flow rate (30 to 60 mL/min) and the total flow rate (300 to 400 mL/min) on the cut-off size D50 (particle diameter with 50% counting efficiency) was investigated using response surface design of experiments (DoE). Thirty one test points with varying operational conditions including 7 replicated test points were performed to evaluate the primary effects of operational parameters and their interactive effects on the cut-off size D50 of the IUCPC. The results indicated that the saturator temperature and the condenser temperature had a greater impact on the cut-off size D50 than the capillary flow rate and the total flow rate. CFD simulation demonstrated that high capillary flow rate in the IUCPC could induce a wider diffusion zone for the core aerosol flow, making more particles enter low supersaturation area near the wall. According to the ANOVA (analysis of variance), the most influential factor for the cut-off size D50 was the condenser temperature, followed by the saturator temperature, the total flow rate and the capillary flow rate. Some interactive effects from combinations of factors on the D50 are also significant. R-squared value of 0.9176 implied that the ultimate model for the IUCPC D50 consistently fits the experimental results.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Aerosol Science - Volume 106, April 2017, Pages 11-23
نویسندگان
, , , , , ,