کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
7234606 1470991 2013 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Development of a microcantilever-based immunosensing method for mycotoxin detection
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Development of a microcantilever-based immunosensing method for mycotoxin detection
چکیده انگلیسی
We here report on the development of an innovative immunosensing method for mycotoxin detection, based on antibody-immobilized microcantilever resonators, a promising label free biosensing technique. A considerable part of the work is devoted to show the effect on microcantilever resonance frequency of the composition of the incubation buffer, as well as of the washing and drying procedure. We show the feasibility of using microcantilever resonator arrays to effectively identify total aflatoxins and ochratoxin A, at low concentrations (3 ng/mL and less than 6 ng/mL, respectively), with relatively low uncertainty (about 10%) and good reproducibility for the same target concentration. Furthermore, the developed immunosensing method shows a limited cross-reactivity to different mycotoxins, paving the way to a highly specific technique, able to identify different mycotoxins in the sample. To our knowledge, this work represents the first example in literature of successfully immunodetection of low concentrations of multiple mycotoxins by microcantilever resonator arrays.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Biosensors and Bioelectronics - Volume 40, Issue 1, 15 February 2013, Pages 233-239
نویسندگان
, , , , , , , , , , ,