عنوان انگلیسی مقاله ISI
Effect of injection matrix concentration on peak shape and separation efficiency in ion chromatography
ترجمه فارسی عنوان
اثر غلظت ماتریس تزریق در شکل پیک و بازده جداسازی در کروماتوگرافی یونی
کلمات کلیدی
ستون تک ستونی، گسترش قله، اثر یون ماتریس، ارزیابی حساسیت تزریق، تزریق،
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
چکیده انگلیسی
In HPLC, injection of solvents that differ from the eluent can result in peak broadening due to solvent strength mismatch or viscous fingering. Broadened, distorted or even split analyte peaks may result. Past studies of this injection-induced peak distortion in reversed phase (RPLC) and hydrophilic interaction (HILIC) liquid chromatography have led to the conclusion that the sample should be injected in the eluent or a weaker solvent. However, there have been no studies of injection-induced peak distortion in ion chromatography (IC). To address this, injection-induced effects were studied for six inorganic anions (F−, Cl−, NO2−, Br−, NO3− and SO42−) on a Dionex AS23 IC column using a HCO3−/CO32− eluent. The VanMiddlesworth-Dorsey injection sensitivity parameter (s) showed that IC of anions has much greater tolerance to the injection matrix (HCO3−/CO32− herein) mismatch than RPLC or HILIC. Even when the injection contained a ten-fold greater concentration of HCO3−/CO32− than the eluent, the peak shapes and separation efficiencies of six analyte ions did not change significantly. At more than ten-fold greater matrix concentrations, analyte anions that elute near the system peak of the matrix were distorted, and in the extreme cases exhibited a small secondary peak on the analyte peak front. These studies better guide the degree of dilution needed prior to IC analysis of anions.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Chromatography A - Volume 1371, 5 December 2014, Pages 177-183
نویسندگان
, ,