کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
774072 1462992 2013 10 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Metallurgical failure analysis of acetabular metal-backed screws
موضوعات مرتبط
مهندسی و علوم پایه سایر رشته های مهندسی مهندسی صنعتی و تولید
پیش نمایش صفحه اول مقاله
Metallurgical failure analysis of acetabular metal-backed screws
چکیده انگلیسی


• We analyzed metal backed screw fractures in two cases of premature total hip arthroplasty revision.
• We found microstructural factors associated to failure of the metal backed screws.
• The performance of the metal backed screws can be enhanced from improved microstructure.

Screws are often used in total hip arthroplasty to increase the cementless acetabular metal-backed fixation stability. This study aimed at identifying metallurgical factors related to metal-backed screw fractures in two cases of premature total hip arthroplasty revision. A preliminary analysis of radiographs and retrieved parts was performed. Fractured screws were subjected to fracture analysis, energy dispersive X-ray spectroscopy (EDS) and metallographic analysis. A finite element analysis was performed to predict the stresses experienced by the screws. Two factors were identified: titanium alloy microstructure, consisting of alpha phase plates with oxygen-rich alpha case and as cast Cr–Co alloy with dendritic microstructure and unfavorably aligned shrinkage formation. Finite element analysis has showed enough level of stress to predict screw fracture in the case of unbounded metal-backed and bone interface.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Engineering Failure Analysis - Volume 32, September 2013, Pages 178–187
نویسندگان
, , , , ,