کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
8249198 1533352 2017 9 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Compass model-based quality assurance for stereotactic VMAT treatment plans
کلمات کلیدی
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Compass model-based quality assurance for stereotactic VMAT treatment plans
چکیده انگلیسی
Compass is suitable for QA of SBRT and SRT treatment plans. However, the QA process should include wide set of DVH-based dose parameters and 3D gamma analysis should be the preferred method when performing clinical patient QA. The results suggest that the Compass should not be used for smaller field sizes than 3 × 3 cm2 or the beam model should be adjusted separately for both small (FS ≤ 3 cm) and large (FS > 3 cm) field sizes.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Physica Medica - Volume 44, December 2017, Pages 42-50
نویسندگان
, , ,