A diagnostic curiosity of isolated androstenedione elevation due to autoantibodies against horseradish peroxidase label of the immunoassay
ترجمه فارسی عنوان
یک کنجکاوی تشخیصی از افزایش اندروستندین جدا شده به علت آنتی بادی های ضد سرطان پراکسیداز در آزمایش ایمنی
موضوعات مرتبط
علوم زیستی و بیوفناوری بیوشیمی، ژنتیک و زیست شناسی مولکولی زیست شیمی
چکیده انگلیسی
Two sisters with hirsutism presented with mild hirsutism and isolated, grossly elevated (> 34.9 nmol/L) serum concentrations of androstenedione measured by competitive, homogeneous immunoassay. The clinically discordant laboratory results prompted us to look for assay interference. In this immunoassay, horseradish peroxidase (HRP)-conjugated androstenedione competes with endogenous androstenedione for binding with the solid-phase polyclonal rabbit IgG antibodies. After a wash step, the amount of signal generated by the bound HRP conjugate is inversely proportional to the androstenedione concentration. Alternative analysis by tandem mass spectrometry (a good first line option for troubleshooting) and repeating the competitive immunoassay after polyethylene glycol treatment returned androstenedione concentrations within reference limits. These findings suggested that the original result was spuriously elevated due to assay interference. Additionally, the patient samples were pre-incubated with heterophile blocking reagents, normal rabbit IgG antibodies and HRP-conjugated normal goat IgG antibodies, followed by repeat measurement using the immunoassay. Only samples pre-incubated with HRP-conjugate returned significantly lower androstenedione (9.5 and 12.5 nmol/L, respectively), implying neutralisation of the interfering antibodies. Androstenedione remained grossly elevated in the other experiments. This deductive exercise showed that the interference is due to autoantibodies against the HRP label used in the immunoassay. Another immunoassay using HRP label (5α-dihydrotestosterone) also produced gross elevation that was normal by tandem mass spectrometry analysis. Assay interferences, though not uncommon, are frequently overlooked. Laboratory results discordant with clinical features should prompt consideration of assay interference to avoid unnecessary investigations and treatment. This is the first report of autoantibodies against the HRP label used in immunoassay.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Clinica Chimica Acta - Volume 476, January 2018, Pages 103-106
نویسندگان
, , , , , , , , ,