کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
868412 909808 2010 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Nanoparticle-based biosensor for the detection of emerging pandemic influenza strains
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Nanoparticle-based biosensor for the detection of emerging pandemic influenza strains
چکیده انگلیسی

Electrically active magnetic (EAM) nanoparticles, consisting of aniline monomer polymerized around gamma iron(III) oxide (γ-Fe2O3) cores, serve as the basis of a direct-charge transfer biosensor developed for detection of surface glycoprotein hemagglutinin (HA) from the Influenza A virus (FLUAV) H5N1 (A/Vietnam/1203/04). H5N1 preferentially binds α2,3-linked host glycan receptors. EAM nanoparticles were immunofunctionalized with antibodies against target HA. Glycans preincubated with HA in 10% mouse serum were incubated with anti-HA–EAM complexes. The anti-HA–EAM complexes effectively acted as immunomagnetic separator of HA from mouse serum matrix. EAM nanoparticles served as the biosensor transducer for cyclic voltammetry measurements. The polyaniline was made electrically active by hydrochloric acid doping. Experimental results indicate that the biosensor is able to detect recombinant H5 HA at 1.4 μM in 10% mouse serum, with high specificity for H5 as compared to H1 (H1N1 A/South Carolina/1/18). This novel design applies EAM nanoparticles in a sensitive, specific, affordable, and easy-to-use biosensor with applications in disease monitoring and biosecurity.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Biosensors and Bioelectronics - Volume 26, Issue 4, 15 December 2010, Pages 1346–1352