کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
869566 909833 2009 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Determination of ascorbic acid by polypyrrole potentiometric detector in ion chromatography
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Determination of ascorbic acid by polypyrrole potentiometric detector in ion chromatography
چکیده انگلیسی

A potentiometric detector, based on electrodeposition of polypyrrole onto Pt electrode, was investigated in this study. The chromatographic performance of the PPy detector was investigated in ion chromatography. The resulting PPy detector exhibited good performance for AA determination with a wide linear range (1.0 × 10−6 to 1.0 × 10−2 M), a highly reproducible response (R.S.D. of 2.45%), without any interference and long-term stability. The calculated detection limit was 7.0 × 10−5 mM at 3σ. The calibration results showed good Nernstian behavior and also satisfactory low detection limit. In order to verify the reliability of the PPy detector, it was applied to the determination of AA in pharmaceutical samples. The results were satisfactory and agreed closely with the manufacturers’ stated contents. The PPy electrode can be used as an alternative novel potentiometric detector material for determination of AA in standards and pharmaceutical samples.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Biosensors and Bioelectronics - Volume 24, Issue 12, 15 August 2009, Pages 3492–3497
نویسندگان
, , , ,