کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
875631 | 910789 | 2016 | 6 صفحه PDF | دانلود رایگان |
• The study developed a modified MUNIX technique and evaluated its performance using motoneuron pool and surface EMG models.
• The modified MUNIX can overcome the effects of MUAP amplitude reduction on the standard MUNIX.
• However, the modified MUNIX is less sensitive than the standard MUNIX for tracking motor unit number changes.
The motor unit number index (MUNIX) technique has provided a quick and convenient approach to estimating motor unit population changes in a muscle. Reduction in motor unit action potential (MUAP) amplitude can lead to underestimation of motor unit numbers using the standard MUNIX technique. This study aims to overcome this limitation by developing a modified MUNIX (mMUNIX) technique. The mMUNIX uses a variable that is associated with the area of compound muscle action potential (CMAP) rather than an arbitrary fixed value (20 mV ms) as used in the standard MUNIX to define the output. The performance of the mMUNIX was evaluated using motoneuron pool and surface electromyography (EMG) models. With a fixed motor unit number, the mMUNIX output remained relatively constant with varying degrees of MUAP amplitude changes, while the standard MUNIX substantially underestimated the motor unit number in such cases. However, when MUAP amplitude remained unchanged, the mMUNIX showed less sensitivity than the standard MUNIX in tracking motor unit loss. The current simulation study demonstrated both the advantages and limitations of the standard and modified MUNIX techniques, which can help guide appropriate application and interpretation of MUNIX measurements.
Journal: Medical Engineering & Physics - Volume 38, Issue 2, February 2016, Pages 115–120