کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
9743491 1491192 2005 8 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
A new method and apparatus for on-site detection of trace levels of chemical warfare agents
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
A new method and apparatus for on-site detection of trace levels of chemical warfare agents
چکیده انگلیسی
A pre-concentration/fast gas chromatograph for the sampling of toxic vapors from air was developed. Pre-concentration is performed by filtering the contaminated air through a tube containing absorbing material. The sample is thermally desorbed and separated on a short chromatographic column using air as a carrier gas. Detection is performed with a commercial hand-held chemical warfare agent detector. The effects of the airflow rate, sample volume and system temperatures on the analytical performance were studied. Detection of organophosphorus vapors at concentrations in the range 4 × 10−4 to 4 × 10−5 mg/m3 was achieved within 2.5 min. The ability to couple a portable detector to a pre-concentration/fast GC using air as a carrier gas encourages the development of a hand-held device for real-time on-site monitoring of low levels of CWA in the air.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Analytica Chimica Acta - Volume 548, Issues 1–2, 29 August 2005, Pages 58-65
نویسندگان
, , , , ,