کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
9748928 1493796 2005 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Integrated capillary electrophoresis amperometric detection microchip with replaceable microdisk working electrode
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Integrated capillary electrophoresis amperometric detection microchip with replaceable microdisk working electrode
چکیده انگلیسی
The interference of separation high voltage with the electrochemical detection is a major challenge to the microchip capillary electrophoresis-electrochemical detection systems with end-channel detection mode. Using dopamine and catechol as model analytes, the influences of channel cross-sectional area and channel-to-electrode distance on the high-voltage interference, accordingly on the separation and detection performances of the microchip capillary electrophoresis-electrochemical detection system were investigated. With the increase of the channel cross-sectional area from 312 through 450-615 μm2, the apparent half-wave potentials of hydrodynamic voltammetry for dopamine at the field strength of 288 V/cm shifted positively from 285 through 330-400 mV. By using a chip with the smallest channel cross-section (312 μm2 with top width of 37.3 μm and depth of 8.9 μm) the residual high-voltage field in the detection cell was small, so that detection was conducted at a channel-to-electrode distance of 20 μm to achieve better performances of separation and detection.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Chromatography A - Volume 1080, Issue 2, 8 July 2005, Pages 192-198