کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
9875864 1533331 2005 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Implementation of enhanced dynamic wedges in pinnacle treatment planning system
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Implementation of enhanced dynamic wedges in pinnacle treatment planning system
چکیده انگلیسی
Enhanced dynamic wedges (EDW) provide many advantages over traditional hard wedges for linear accelerator treatments. Along with these advantages comes the responsibility of ensuring that this complex technology delivers the correct dose to patients. This involves determining the enhanced dynamic wedge factors for various field sizes and depths for use in the hand calculation of monitor units (MUs). The accurate representation of dynamic wedges in the treatment planning computer must also be ensured. This is required so that the final isodose distributions are correct and the MUs calculated by the treatment planning computer match those determined by hand calculation. We have commissioned and implemented the use of EDW in the Pinnacle radiation therapy planning system. The modeled dose profiles agree with the measured ones with a maximum difference of 2%. The MUs generated by Pinnacle are also within 2% of those calculated independently. The process of data collection and verification, beam modeling, and a discussion of a potential pitfall encountered in this process are presented in this paper.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Medical Dosimetry - Volume 30, Issue 4, Winter 2005, Pages 228-232
نویسندگان
, , ,