کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
997 70 2011 8 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Selection of the implant and coating materials for optimized performance by means of nanoindentation
موضوعات مرتبط
مهندسی و علوم پایه مهندسی شیمی بیو مهندسی (مهندسی زیستی)
پیش نمایش صفحه اول مقاله
Selection of the implant and coating materials for optimized performance by means of nanoindentation
چکیده انگلیسی

Mechanical compatibility between a coating and a substrate is important for the longevity of implant materials. While previous studies have utilized the entire coating for analysis of mechanical compatibility of the surface, this study focuses on the nanoindentation of a uniformly thermally sprayed splat. Hydroxyapatite was thermally sprayed to create a homogeneous deposit density, as confirmed by microRaman spectroscopy, of amorphous calcium phosphate. Substrates were commercially pure Ti, Ti–6Al–4V, Co–Cr alloy and stainless steel. Nanoindentation revealed that splats deposited on the different metals have similar hardness and elastic modulus values of 4.2 ± 0.2 GPa and 80 ± 3 GPa, respectively. The mechanical properties were affected by the substrate type more than residual stresses, which were found to be low. It is recommended that amorphous calcium phosphate is annealed to relieve the quenching stress or that appropriate temperature histories are chosen to relax the stress created in cooling the coating assembly.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Acta Biomaterialia - Volume 7, Issue 2, February 2011, Pages 874–881