Validation of a screening method based on liquid chromatography coupled to high-resolution mass spectrometry for analysis of perfluoroalkylated substances in biota
چکیده انگلیسی
A screening method for analysis of perfluoroalkylated substances (PFAS) in biota samples has been developed and validated using liver samples from polar cod (Boreogadus saida) and glaucous gull (Larus hyperboreus). The method was based on extraction of target compounds from homogenised samples into the solvent mixture used as mobile phase in high-performance liquid chromatography (HPLC), i.e. methanol/water (50:50; 2 mM ammonium acetate). The extract was filtered and directly injected into a HPLC/time-of-flight mass spectrometry (TOF-MS) system. Quantification was performed using 7H-perfluoroheptanoic acid as internal standard and a calibration standard solution dissolved in sample extract for each matrix type (matrix-matched calibration standard). The method is very time and cost efficient. Except for long-chain compounds and perfluorooctane sulfonamide (which cannot be covered by this method), recoveries were between 60% and 115% and method detection limits were in the range 0.04-1.3 ng/g wet weight. Blank values could be neglected with the exception of perfluorooctane sulfonate (PFOS), perfluorohexanoic acid (PFHxA) and perfluorooctanoic acid (PFOA). One of the major challenges in PFAS analysis is ionisation disturbance by co-eluting matrix in the ion source of the mass spectrometer. Both matrix and analyte specific signal enhancement and suppression was observed and quantified. Repeated extractions (n = 3) gave relative standard deviations (RSD) <35% for all PFAS. Accuracy was examined by comparing the screening method to the generally applied ion pair extraction (IPE) method. PFAS concentration values of a glaucous gull liver sample deviated by less than 30% for the two methods, provided that matrix-matched calibration standards were employed in both methods.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Chromatography A - Volume 1081, Issue 2, 22 July 2005, Pages 210-217
نویسندگان
, ,