کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
10553645 967888 2009 10 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Characterization and resolution of reversed phase HPLC chromatography failure attributed to sulfobutylether-β-cyclodextrin in a pharmaceutical sample preparation
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Characterization and resolution of reversed phase HPLC chromatography failure attributed to sulfobutylether-β-cyclodextrin in a pharmaceutical sample preparation
چکیده انگلیسی
A reversed phase HPLC method developed for a drug product formulation using hydroxypropyl-β-cyclodextrin (HPCD) was rendered ineffective for analyzing a similar formulation containing sulfobutylether-β-cyclodextrin (SBECD). The active pharmaceutical ingredient (API) and the majority of its impurities became more strongly retained, eluting as an incoherent conglomerate of peaks. Furthermore, this phenomenon was reproduced in subsequent injections of the API reference standard. Based on HPLC and LC-ESI-MS studies, the chromatography failure was attributed to the accumulation of SBECD on the HPLC column. The subsequent interaction of the API with bound SBECD resulted in the aberrant chromatography. An anion-exchange solid-phase extraction treatment was developed and qualified to selectively remove SBECD from sample solutions, thereby allowing the same HPLC method to be used. The sample treatment procedure exhibited suitable accuracy and precision for quantitating the API and its impurities, and resulted in typical chromatographic profiles.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Pharmaceutical and Biomedical Analysis - Volume 49, Issue 3, 5 April 2009, Pages 660-669
نویسندگان
, , ,