Simple automated generation of gradient elution conditions in sequential injection chromatography using monolithic column
چکیده انگلیسی
The GE-SIC separation was performed with a FIAlab® 3000 SIC set-up (USA) equipped with an Onyx™ Monolithic C18 (25 mm × 4.6 mm, Phenomenex®) column, a six-port selection valve, a 5-mL syringe pump and a fiber-optics UV CCD detector. Ketoprofen was used as an internal standard (IS). The gradient elution was achieved by automated reproducible mixing of acetonitrile and aqueous 0.2% phosphoric acid in the holding coil of the SIC system. Different profiles of the gradient elution were tested. The optimal gradient using two mobile phases 30:70 and 50:50 of acetonitrile/0.2% phosphoric acid (v/v) was achieved under the optimum flow rate 1.2 mL min−1. The chromatographic resolution R between the peaks of all solutes (including the IS) was >2.00. The repeatability of retention times was characterized by the RSD values 0.18-0.30% (n = 6). Net separation time was 3.5 min and the mobile phase consumption was 4.5 mL for a single GE-SIC assay. The figures of merit of the novel GE-SIC method compared well with those of conventional HPLC.